12E5119 Instruction Manual QuickTOC
®
ultra
IECEx 277
13 Operating, Care and Maintenance Logs
13.12 Care and Maintenance Log (CO2-Remover)
LAR | PROCESS ANALYSERS AG
13.12 Care and Maintenance Log (CO
2
-Remover)
Interval Action
Kind of
Action
OK Comment
12,000
operating
hours
(or every 2
years)
Replace the desiccant cartridges Maintenance
Replace the internal ball valves Maintenance
Replace all seals which removed
during maintenance
Maintenance
24,000
operating
hours
(or every 4
years)
Replace the exhaust valves Maintenance
Date: Signature: