4 Installation
4.2 Site Selection - Ambient Conditions LAR | PROCESS ANALYSERS AG
4.2 Site Selection - Ambient Conditions
The following ambient conditions apply for choosing the analyser location:
Dry and frost-proof
Allowed temperature ranget 5°C - 35°C
Max. relative humidity 80%
No direct exposure to sunlight
No aggressive environment for housing type IP54
Mains voltage 115/230 V at 50/60 Hz, 16 A fusing (K-characteristics)
Warning
Warning of incorrect electric connections
The analyser may be damaged if it is not connected to the mains matching the
data stated on the nameplate.
Before the analyser is switched on, a check must be run on whether the local
voltage supply matches that on the nameplate.
If this power supply is not available, please contact your local partner or the
Technical Support of LAR (Chapter 15 on page 283).
Do not switch the analyser on before using it for the first time without the pre-
sence of a technician authorised by LAR - otherwise this voids the warranty of
your analyser.
!