8 Care and Maintenance
8.3 Performing the Visual Inspections
LAR | PROCESS ANALYSERS AG
Technical Support (Chapter 15.1 from page 283).
3. If the criterion is fulfilled, proceed to the next step.
8.3.3 Injection System
1. Ensure that the analyser is in offline mode.
2. Check the injection system for air bubbles in the tube or glass syringe.
3. Rinse the injection system if there are air bubbles in the tube or glass syringe (Chapter 8.5.1 on
page 172).
4. Once the test criterion is met, proceed with the next step.