226 Instruction Manual QuickTOC
®
ultra
IECEx 12E5119
9 Accessories and Options
9.4 Overflow Sampler LAR | PROCESS ANALYSERS AG
9.4.3 Set-Up
1. Install the LAR overflow sampler on the wall immediately to the right of the analyser. In this case,
the overflow sampler should be mounted so that the sample inlet of the analyser is approximately
at the same level as the sample outlet (4, Fig. 144, page 225) of the overflow sampler.
2. Set the sample purging time in the software under "Service actions" (see Chapter 7.8.2 from
page 118) in the menu item "Rinse sample tubes" to a minimum rinse time of 30 seconds.
1. Connect the inlet (2) with the sample tube.
2. Connect the sample outlet (4) of the overflow sampler with the sample inlet of the analyser.
3. Open the sample supply.
Abb. 145: Setting the purge time
30