218 Instruction Manual QuickTOC
®
ultra
IECEx 12E5119
9 Accessories and Options
9.3 Proceed analogously to the replacement of the soda lime in Chapter 9.2.7.1 from page 213FlowSampler® LAR | PRO-
9.3 Proceed analogously to the replacement of the soda lime in Chapter 9.2.7.1 from
page 213FlowSampler
®
The FlowSampler
®
is a sample feed, which works without filtration according to the principle of inertia
separation because the sample is extracted in the centre of the sample flow in the opposite direction to
the main flow of the waste water. This means all solid and heavy coarse materials (such as sand) are
reliably removed. Smaller solid particles are caught, whereby the sample enters the analyser in its pure
form. The FlowSampler
®
can also master the most difficult challenges - including sample taking at a clar-
ification plant inlet in front of the coarse screen.
9.3.1 Variants of FlowSampler
®
LAR offers two variants which differ by diameter size:
• FlowSampler
®
DN25
• FlowSampler
®
DN17
Material (depends on the temperature of the sample):
• from 00°C to 50°C: Polyvinylchloride (PVC)
• from 50°C to 90°C: Polypropylene (PP)
Number of sample streams:
• The FlowSampler
®
can be used for a maximum of six sample streams.
Notice
The diameter size is dependent on the realisable flow rate of the sample.
If you have any question, please contact the Sales Departement of LAR
(Chapter 15 on page 283).
i