5 Start-up
5.10 Installation of the Pump Tubes
LAR | PROCESS ANALYSERS AG
Proceed as follows:
1. Locate all sample pumps by using Tab. 2, page 21.
2. Open the upper lid (1) of the sample pump.
3. Place the tube (2) close to the roller (3).
4. Pay attention not to twist the tube (2).
5. Pull the tube (2) down on both sides to make sure that the tube (2) is in both sides of the guide (4).
6. Select the value according to the diameter of the used tubing (2) by means of the adjustment
wheels (5). The value can bei either 3,2 mm or 4,8 mm.
7. Close the upper lid (1) of the sample pump.
8. Proceed analogously with the other pump tubings and sample pumps.
Fig. 47: Installation of the Tubes into the Sample Pump