Software Version C
- 137 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
Note!
In case of using several sensors, the system zero point can be set
independently!
10. After setting the system zero point, all measuring results are 0.00 or 100 %
for transmission measurements (format and unit correspond to the set
product).
If linearization table or slope + offset functions are used to change the zero
point the calculated value will be displayed.
11. Press [CLEAR] twice to return to the main menu.
Note!
Zeroing is carried out for optek inputs, which are used for the measuring result
marked with YES!
The system displays the input assignments of all products in the MEASURING
RESULT MONITOR (see chapter 9.11.8.2.3, page 155)!
Zeroing is not possible for functions using the factory zero point.