Spectrum Measurements
R&S
®
ESR
318User Manual 1175.7068.02 ─ 12
Display range:
DISP:WIND:TRAC:Y 50DB, see DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
on page 844
Range Log 10 dB ← x-Axis Range ← Scaling ← APD
Sets the level display range to 10 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG, see DISPlay[:WINDow<n>]:TRACe<t>:Y:
SPACing on page 668
Display range:
DISP:WIND:TRAC:Y 10DB, see DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
on page 844
Range Log 5 dB ← x-Axis Range ← Scaling ← APD
Sets the level display range to 5 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG, see DISPlay[:WINDow<n>]:TRACe<t>:Y:
SPACing on page 668
Display range:
DISP:WIND:TRAC:Y 5DB, see DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
on page 844
Range Log 1 dB ← x-Axis Range ← Scaling ← APD
Sets the level display range to 1 dB.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG, see DISPlay[:WINDow<n>]:TRACe<t>:Y:
SPACing on page 668
Display range:
DISP:WIND:TRAC:Y 1DB, see DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
on page 844
Range Log Manual ← x-Axis Range ← Scaling ← APD
Opens an edit dialog box to define the display range of a logarithmic level axis man-
ually.
Remote command:
Logarithmic scaling:
DISP:WIND:TRAC:Y:SPAC LOG, see DISPlay[:WINDow<n>]:TRACe<t>:Y:
SPACing on page 668
Display range:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe] on page 844
Range Linear % ← x-Axis Range ← Scaling ← APD
Selects linear scaling for the level axis in %.
The grid is divided into decadal sections.
Measurements