EasyManuals Logo

R&S ESR Series User Manual

R&S ESR Series
1126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #349 background imageLoading...
Page #349 background image
Spectrum Measurements
R&S
®
ESR
346User Manual 1175.7068.02 ─ 12
0.5*Span (span > 0) ← CF Stepsize
Sets the step size for the center frequency to 50 % of the span.
Remote command:
FREQ:CENT:STEP:LINK SPAN, see [SENSe:]FREQuency:CENTer:STEP:LINK
on page 823
FREQ:CENT:STEP:LINK:FACT 50PCT, see [SENSe:]FREQuency:CENTer:STEP:
LINK:FACTor on page 824
0.5*RBW (span > 0) ← CF Stepsize
Sets the step size for the center frequency to 50 % of the resolution bandwidth.
Remote command:
FREQ:CENT:STEP:LINK RBW, see [SENSe:]FREQuency:CENTer:STEP:LINK
on page 823
FREQ:CENT:STEP:LINK:FACT 50PCT, see [SENSe:]FREQuency:CENTer:STEP:
LINK:FACTor on page 824
x*Span (span > 0) ← CF Stepsize
Opens an edit dialog box to set the step size for the center frequency as a percentage
(%) of the span.
Remote command:
FREQ:CENT:STEP:LINK SPAN, see [SENSe:]FREQuency:CENTer:STEP:LINK
on page 823
FREQ:CENT:STEP:LINK:FACT 20PCT, see [SENSe:]FREQuency:CENTer:STEP:
LINK on page 823
x*RBW (span > 0) ← CF Stepsize
Opens an edit dialog box to set the step size for the center frequency as a percentage
(%) of the resolution bandwidth. Values between 1 % and 100 % in steps of 1 % are
allowed. The default setting is 10 %.
Remote command:
FREQ:CENT:STEP:LINK RBW, see [SENSe:]FREQuency:CENTer:STEP:LINK
on page 823
FREQ:CENT:STEP:LINK:FACT 20PCT, see [SENSe:]FREQuency:CENTer:STEP:
LINK on page 823
=Center ← CF Stepsize
Sets the step size to the value of the center frequency and removes the coupling of the
step size to span or resolution bandwidth.
This function is especially useful for measurements of the signal harmonics. In this
case, each stroke of the arrow key selects the center frequency of another harmonic.
=Marker ← CF Stepsize
Sets the step size to the value of the current marker and removes the coupling of the
step size to span or resolution bandwidth.
This function is especially useful for measurements of the signal harmonics. In this
case, each stroke of the arrow key selects the center frequency of another harmonic.