Remote Control
R&S
®
ESR
824User Manual 1175.7068.02 ─ 12
[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor <Factor>
This command defines a step size factor if the center frequency step size is coupled to
the span or the resolution bandwidth.
Parameters:
<Factor> Range: 1 to 100
*RST: 10
Default unit: PCT
Example:
FREQ:CENT:STEP:LINK:FACT 20PCT
Manual operation: See "0.1*Span (span > 0)" on page 345
See "0.1*RBW (span > 0)" on page 345
See "0.5*Span (span > 0)" on page 346
See "0.5*RBW (span > 0)" on page 346
[SENSe:]FREQuency:MODE <Mode>
This command selects the frequency mode.
between span > 0 and zero span in the "Spectrum" mode.
Parameters:
<Mode> CW
Selects the time domain (span = 0).
In the time domain, define the measurement frequency with
[SENSe:]FREQuency:CENTer.
FIXed
Selects the time domain (span = 0).
In the time domain, define the measurement frequency with
[SENSe:]FREQuency:CENTer.
SWEep
Selects the frequency domain (span > 0).
In the frequency domain, define the frequency characteristics
with the following commands:
• [SENSe:]FREQuency:STARt
• [SENSe:]FREQuency:STOP
• [SENSe:]FREQuency:SPAN
• [SENSe:]FREQuency:CENTer
*RST:
SWEep
Example:
FREQ:MODE SWE
Selects the frequency domain.
Usage: SCPI confirmed
[SENSe:]FREQuency:OFFSet <Offset>
This command defines the frequency offset.
Remote Commands in Spectrum Analyzer Mode