Spectrum Measurements
R&S
®
ESR
375User Manual 1175.7068.02 ─ 12
with a special option, model or (measurement) mode, this information is provided in the
corresponding softkey description.
Trg/Gate Source..........................................................................................................375
└ Free Run.......................................................................................................375
└ External.........................................................................................................375
└ Video.............................................................................................................376
└ RF Power......................................................................................................376
└ IF Power/BB Power.......................................................................................376
└ Time.............................................................................................................. 377
Trg/Gate Level.............................................................................................................377
Trg/Gate Polarity......................................................................................................... 377
Trigger Offset.............................................................................................................. 378
Repetition Interval....................................................................................................... 378
Trigger Hysteresis....................................................................................................... 378
Trigger Holdoff.............................................................................................................378
Gated Trigger.............................................................................................................. 378
Gate Settings.............................................................................................................. 379
└ Gate Mode (Lvl/Edge)...................................................................................379
└ Gate Delay....................................................................................................379
└ Gate Length (Gate Mode Edge)....................................................................380
└ Trg/Gate Source........................................................................................... 380
└ Trg/Gate Level.............................................................................................. 380
└ Trg/Gate Polarity...........................................................................................380
└ Sweep Time.................................................................................................. 380
└ Res BW Manual............................................................................................380
Trg/Gate Source
Opens the "Trigger/Gate Source" dialog box to select the trigger/gate mode.
As gate modes, all modes except "Power Sensor" are available. For details see also
Chapter 6.2.8.3, "Using Gated Sweep Operation", on page 381.
The default setting is "Free Run". If a trigger mode other than "Free Run" has been set,
the enhancement label "TRG" is displayed and the trigger source is indicated.
Note: When triggering or gating is activated, the squelch funciton is automatically dis-
abled (see "Squelch" on page 421).
Remote command:
TRIGger<n>[:SEQuence]:SOURce on page 839
[SENSe:]SWEep:EGATe:SOURce on page 842
Free Run ← Trg/Gate Source
The start of a sweep is not triggered. Once a measurement is completed, another is
started immediately.
Remote command:
TRIG:SOUR IMM, see TRIGger<n>[:SEQuence]:SOURce on page 839
External ← Trg/Gate Source
Defines triggering via a TTL signal at the "EXT TRIG/GATE IN" input connector on the
rear panel.
Configuration