Remote Control
R&S
®
ESR
699User Manual 1175.7068.02 ─ 12
CALCulate<n>:DELTamarker<m>:MAXimum[:PEAK]
This command positions the delta marker to the current trace maximum. If necessary,
the corresponding delta marker is activated first.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:DELT3:MAX
Sets delta marker 3 to the maximum value of the associated
trace.
Manual operation: See "Peak" on page 205
CALCulate<n>:DELTamarker<m>:MAXimum:RIGHt
This command positions the delta marker to the next smaller trace maximum on the
right of the current value (i.e. ascending X values). The corresponding delta marker is
activated first, if necessary.
If no next higher minimum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:DELT:MAX:RIGH
Sets delta marker 1 to the next smaller maximum value to the
right of the current value.
Manual operation: See "Next Mode" on page 205
CALCulate<n>:DELTamarker<m>:MINimum:LEFT
This command positions the delta marker to the next higher trace minimum on the left
of the current value (i.e. descending X values). The corresponding delta marker is acti-
vated first, if necessary.
If no next higher minimum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: -200) is produced.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Example:
CALC:DELT:MIN:LEFT
Sets delta marker 1 to the next higher minimum to the left of the
current value.
Manual operation: See "Next Mode" on page 205
Remote Commands in Receiver Mode