EasyManua.ls Logo

R&S ESR Series

R&S ESR Series
1126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Receiver Mode
R&S
®
ESR
205User Manual 1175.7068.02 ─ 12
"Specifying the Search Limits" on page 413
"Specifying the Search Range" on page 413
"Examining a Signal at the Center in Detail" on page 413
"Specifying the Suitable Peak Excursion" on page 414
Select Marker (No)
Opens a submenu to select one of 16 markers and define whether the marker is a nor-
mal or a delta marker (see "Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
Norm/Delta" on page 201). "(No)" indicates the number of the currently active marker.
Peak
Sets the active marker/delta marker to the highest maximum of the trace.
Remote command:
CALCulate<n>:MARKer<m>:MAXimum[:PEAK] on page 692
CALCulate<n>:DELTamarker<m>:MAXimum[:PEAK] on page 699
Next Peak
Sets the active marker/delta marker to the next maximum of the selected trace.
Remote command:
CALCulate<n>:MARKer<m>:MAXimum:NEXT on page 692
CALCulate<n>:DELTamarker<m>:MAXimum:NEXT on page 698
Min
Sets the active marker/delta marker to the minimum of the selected trace.
Remote command:
CALCulate<n>:MARKer<m>:MINimum[:PEAK] on page 694
CALCulate<n>:DELTamarker<m>:MINimum[:PEAK] on page 700
Next Min
Sets the active marker/delta marker to the next minimum of the selected trace.
Remote command:
CALCulate<n>:MARKer<m>:MINimum:NEXT on page 693
CALCulate<n>:DELTamarker<m>:MINimum:NEXT on page 700
Next Mode
Selects the mode of the Next Peak or Next Min softkey.
Three settings are available:
"<"
Sets the active marker/delta marker to the next maximum/minimum
left to the marker of the selected trace.
"abs"
Sets the active marker/delta marker to the next lower maximum/
higher minimum of the selected trace.
">"
Sets the active marker/delta marker to the next maximum/minimum
right to the marker of the selected trace.
Common Analysis Functions

Table of Contents