Remote Control
R&S
®
ESR
768User Manual 1175.7068.02 ─ 12
Suffix:
<range>
.
1...20
range
Parameters:
<Level> -400 to in 400 dBm
*RST: 13 dBm
Example:
ESP:RANG1:LIM:ABS:STOP 20
Sets an absolute limit of 20 dBm at the stop frequency of the
range.
Manual operation: See "Abs Limit Stop" on page 261
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:RELative:STARt <Limit>
This command sets a relative limit value at the start frequency of the specified range.
Different from manual operation, this setting is independent of the defined limit check
type.
Suffix:
<range>
.
1...20
range
Parameters:
<Limit> -400 to in 400 dBc
*RST: -50 dBc
Example:
ESP:RANG3:LIM:REL:STAR -20
Sets a relative limit of -20 dBc at the start frequency of the
range.
Manual operation: See "Rel Limit Start" on page 261
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:RELative:STOP <Limit>
This command sets a relative limit value at the stop frequency of the specified range.
Different from manual operation, this setting is independently of the defined limit check
type.
Suffix:
<range>
.
1...20
range
Parameters:
<Limit> -400 to in 400 dBc
*RST: -50 dBc
Example:
ESP:RANG3:LIM:REL:STOP 20
Sets a relative limit of 20 dBc at the stop frequency of the range.
Manual operation: See "Rel Limit Stop" on page 261
Remote Commands in Spectrum Analyzer Mode