Example:
ESP:RANG3:INP:GAIN:STATe ON
Switches the preamplifier for range 3 on or off.
Manual operation: See "Preamp" on page 260
[SENSe:]ESPectrum:RANGe<range>:INSert <Mode>
This command inserts a new range before or after the specified range. The range num-
bers are updated accordingly.
Suffix:
<range>
.
1...20
range
Parameters:
<Mode> AFTer | BEFore
Example:
ESP:RANG3:INS BEF
Inserts a new range before range 3.
ESP:RANG1:INS AFT
Inserts a new range after range 1.
Manual operation: See "Insert before Range" on page 262
See "Insert after Range" on page 262
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:ABSolute:STARt <Level>
This command sets an absolute limit value at the start frequency of the specified
range. Different from manual operation, this setting is independently of the defined limit
check type.
Suffix:
<range>
.
1...20
range
Parameters:
<Level> -400 to in 400 dBm
*RST: 13 dBm
Example:
ESP:RANG1:LIM:ABS:STAR 10
Sets an absolute limit of 10 dBm at the start frequency of the
range.
Manual operation: See "Abs Limit Start" on page 261
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:ABSolute:STOP <Level>
This command sets an absolute limit value at the stop frequency of the specified
range. Different from manual operation, this setting is independent of the defined limit
check type.