Remote Control
R&S
®
ESR
767User Manual 1175.7068.02 ─ 12
Example:
ESP:RANG3:INP:GAIN:STATe ON
Switches the preamplifier for range 3 on or off.
Manual operation: See "Preamp" on page 260
[SENSe:]ESPectrum:RANGe<range>:INSert <Mode>
This command inserts a new range before or after the specified range. The range num-
bers are updated accordingly.
Suffix:
<range>
.
1...20
range
Parameters:
<Mode> AFTer | BEFore
Example:
ESP:RANG3:INS BEF
Inserts a new range before range 3.
ESP:RANG1:INS AFT
Inserts a new range after range 1.
Manual operation: See "Insert before Range" on page 262
See "Insert after Range" on page 262
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:ABSolute:STARt <Level>
This command sets an absolute limit value at the start frequency of the specified
range. Different from manual operation, this setting is independently of the defined limit
check type.
Suffix:
<range>
.
1...20
range
Parameters:
<Level> -400 to in 400 dBm
*RST: 13 dBm
Example:
ESP:RANG1:LIM:ABS:STAR 10
Sets an absolute limit of 10 dBm at the start frequency of the
range.
Manual operation: See "Abs Limit Start" on page 261
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:ABSolute:STOP <Level>
This command sets an absolute limit value at the stop frequency of the specified
range. Different from manual operation, this setting is independent of the defined limit
check type.
Remote Commands in Spectrum Analyzer Mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals