Spectrum Measurements
R&S
®
ESR
262User Manual 1175.7068.02 ─ 12
This parameter is only available if the limit check is set accordingly (see "Sweep List
dialog box" on page 258).
Remote command:
[SENSe:]ESPectrum:RANGe<range>:LIMit<source>:RELative:STOP
on page 768
Close Sweep List ← Sweep List ← Spectrum Emission Mask
Closes the "Sweep List" dialog box and updates the measurement results.
Insert before Range ← Sweep List ← Spectrum Emission Mask
Inserts a new range to the left of the currently focused range. The range numbers of
the currently focused range and all higher ranges are increased accordingly. The maxi-
mum number of ranges is 20.
For further details refer to "Ranges and Range Settings" on page 269.
Remote command:
ESP:RANG3:INS BEF, see [SENSe:]ESPectrum:RANGe<range>:INSert
on page 767
Insert after Range ← Sweep List ← Spectrum Emission Mask
Inserts a new range to the right of the currently focused range. The range numbers of
all higher ranges are increased accordingly. The maximum number of ranges is 20.
For further details refer to "Ranges and Range Settings" on page 269.
Remote command:
ESP:RANG1:INS AFT, see [SENSe:]ESPectrum:RANGe<range>:INSert
on page 767
Delete Range ← Sweep List ← Spectrum Emission Mask
Deletes the currently focused range, if possible. The range numbers are updated
accordingly. For further details refer to "Ranges and Range Settings" on page 269.
Remote command:
[SENSe:]ESPectrum:RANGe<range>:DELete on page 764
Symmetric Setup ← Sweep List ← Spectrum Emission Mask
If activated, the current sweep list configuration is changed to define a symmetrical
setup regarding the reference range. The number of ranges to the left of the reference
range is reflected to the right, i.e. any missing ranges on the right are inserted, while
superfluous ranges are removed. The values in the ranges to the right of the reference
range are adapted symmetrically to those in the left ranges.
Any changes to the range settings in active "Symmetric Setup" mode lead to symmetri-
cal changes in the other ranges (where possible). In particular, this means:
Inserting ranges: a symmetrical range is inserted on the other side of the reference
range
Deleting ranges: the symmetrical range on the other side of the reference range is
also deleted
Editing range settings: the settings in the symmetrical range are adapted accord-
ingly
Measurements