Spectrum Measurements
R&S
®
ESR
260User Manual 1175.7068.02 ─ 12
VBW ← Sweep List dialog box ← Sweep List ← Spectrum Emission Mask
Sets the VBW value for this range.
Remote command:
[SENSe:]ESPectrum:RANGe<range>:BANDwidth:VIDeo on page 763
Sweep Time Mode ← Sweep List dialog box ← Sweep List ← Spectrum Emission
Mask
Activates or deactivates the auto mode for the sweep time.
Remote command:
[SENSe:]ESPectrum:RANGe<range>:SWEep:TIME:AUTO on page 770
Sweep Time ← Sweep List dialog box ← Sweep List ← Spectrum Emission Mask
Sets the sweep time value for the range.
Remote command:
[SENSe:]ESPectrum:RANGe<range>:SWEep:TIME on page 769
Ref. Level ← Sweep List dialog box ← Sweep List ← Spectrum Emission Mask
Sets the reference level for the range.
Remote command:
[SENSe:]ESPectrum:RANGe<range>:RLEVel on page 769
RF Att. Mode ← Sweep List dialog box ← Sweep List ← Spectrum Emission
Mask
Activates or deactivates the auto mode for RF attenuation.
Remote command:
[SENSe:]ESPectrum:RANGe<range>:INPut:ATTenuation:AUTO on page 766
RF Attenuator ← Sweep List dialog box ← Sweep List ← Spectrum Emission
Mask
Sets the attenuation value for that range.
Remote command:
[SENSe:]ESPectrum:RANGe<range>:INPut:ATTenuation on page 766
Preamp ← Sweep List dialog box ← Sweep List ← Spectrum Emission Mask
Switches the preamplifier on or off.
Remote command:
[SENSe:]ESPectrum:RANGe<range>:INPut:GAIN:STATe on page 766
Transd. Factor ← Sweep List dialog box ← Sweep List ← Spectrum Emission
Mask
Sets a transducer for the specified range. You can only choose a transducer that fulfills
the following conditions:
The transducer overlaps or equals the span of the range.
The x-axis is linear.
The unit is dB.
Remote command:
[SENSe:]ESPectrum:RANGe<range>:TRANsducer on page 770