Remote Control
R&S
®
ESR
908User Manual 1175.7068.02 ─ 12
Parameters:
<Name> '<name>
Example:
CORR:TRAN:GEN 'SMU01'
Creates the transducer file C:\r_s\instr\trd\SMU01.trd.
Manual operation: See "Save As Trd Factor" on page 479
SOURce<n>:EXTernal<generator>:FREQuency <Frequency>
This command defines a fixed source frequency for the external tracking generator.
This command is only valid for the External Tracking Generator option (R&S FSV-B10).
Suffix:
<n>
.
1...4
window
<m> 1 | 2
tracking generator
Parameters:
<Frequency> Source frequency of the external tracking generator.
*RST: RST value
Example:
SOUR:EXT2:FREQ 10MHz
Manual operation: See "(Fixed) Generator Frequency" on page 488
SOURce<n>:EXTernal<generator>:FREQuency:COUPling[:STATe] <State>
This command couples or decouples the analyzer and generator frequencies.
This command is only valid for the External Tracking Generator option (R&S FSV-B10).
Suffix:
<generator>
.
1 | 2
tracking generator
Parameters:
<State> ON | OFF
ON
OFF
The fixed frequency defined using SOURce<n>:
EXTernal<generator>:FREQuency is used as a source fre-
quency.
*RST: OFF
Example:
SOUR:EXT2:FREQ:COUP ON
Manual operation: See "Source Frequency Coupling" on page 488
Remote Commands to Control the Tracking Generator