EasyManua.ls Logo

R&S ESR Series

R&S ESR Series
1126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
ESR
980User Manual 1175.7068.02 ─ 12
Example:
DISP:PSAV:HOLD 30
Manual operation: See "Display Pwr Save (On/Off)" on page 523
DISPlay:PSAVe[:STATe] <State>
This command switches on or off the power-save mode of the display. With the power-
save mode activated the display including backlight is completely switched off after the
elapse of the response time (see DISPlay:PSAVe:HOLDoff command).
This mode is recommended to protect the display, especially if the instrument is oper-
ated exclusively via remote control.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
DISP:PSAVe ON
Switches on the power-save mode.
Manual operation: See "Display Pwr Save (On/Off)" on page 523
DISPlay:SBAR[:STATe] <State>
This command turns the status bar in the display on and off.
Parameters:
<State> ON | OFF
*RST: ON
Example:
DISP:SBAR OFF
Manual operation: See "Status Bar" on page 520
DISPlay:SKEYs[:STATe] <State>
This command turns the softkey bar in the display on and off.
Parameters:
<State> ON | OFF
*RST: ON
Example:
DISP:SKEY OFF
DISPlay:TBAR[STATe] <State>
This command turns the toolbar in the display on or off.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
DISP:TOOL ON