Remote Control
R&S
®
ESR
993User Manual 1175.7068.02 ─ 12
For FSP (=FSV), the standard parameter coupling of the instrument is used. As a
result, in most cases a shorter sweeptime is used than in case of HP.
This command is only available if a HP language is selected using SYSTem:LANGuage
.
Parameters:
<CouplingType> HP | FSP
*RST: FSP
Example:
SYSTem:HPC HP
Usage: SCPI confirmed
Manual operation: See "Coupling (FSx/HP)" on page 517
SYSTem:IFGain:MODE <Mode>
Configures the internal IF gain settings in HP emulation mode due to the application
needs. This setting is only taken into account for resolution bandwidth < 300 kHz and is
only available if a HP language is selected using SYSTem:LANGuage.
Parameters:
<Mode> NORM | PULS
NORM
Optimized for high dynamic range, overload limit is close to ref-
erence level.
PULS
Optimized for pulsed signals, overload limit up to 10 dB above
reference level.
*RST: NORM
Example:
SYST:IFG:MODE PULS
Usage: SCPI confirmed
Manual operation: See "IF Gain (Norm/Puls)" on page 516
SYSTem:TWALking <Mode>
Defines how remote control programs are evaluated, in particular multiple commands
in one command line.
Note that the default setting is NORM; however, the setting defined by this command is
retained after a [preset] (*RST) operation.
Parameters:
<Mode> NORM
SCPI-conform behaviour
The first command is always referenced to the root node. Sub-
sequent commands, however, are referenced to the same tree
level as the previous command in a message unit.
Network Connection

Table of Contents