System Configuration
R&S
®
ESR
517User Manual 1175.7068.02 ─ 12
Coupling (FSx/HP) ← Compatibility Mode ← GPIB ← General Setup
Controls the default coupling ratios in the HP emulation mode for:
Span and resolution bandwidth (Span/RBW) and
Resolution bandwidth and video bandwidth (RBW/VBW)
For FSP(=FSV), the standard parameter coupling of the instrument is used. As a
result, usually a shorter sweeptime is used than for HP.
This softkey is only available if an HP language is selected via the "GPIB Language"
softkey (see "GPIB Language" on page 515).
Remote command:
SYSTem:HPCoupling on page 992
REV String Factory ← Compatibility Mode ← GPIB ← General Setup
Selects the default response to the REV? query for the revision number (HP emulation
only, see Chapter 11.18, "GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B
and 8594E", on page 1039).
Remote command:
SYSTem:REVision:FACTory on page 991
REV String User ← Compatibility Mode ← GPIB ← General Setup
Opens an edit dialog box to enter a user-defined revision number in response to the
REV? query (HP emulation only, see Chapter 11.18, "GPIB Commands of HP Models
856xE, 8566A/B, 8568A/B and 8594E", on page 1039). Max. 36 characters are
allowed.
Remote command:
SYSTem:REVision[:STRing] on page 991
Tree Walking ← Compatibility Mode ← GPIB ← General Setup
Defines how remote control programs are evaluated, in particular multiple commands
in one command line.
Note that this setting is retained after a [preset] or recall operation.
See the SCPI command for an example.
"Normal"
(Initial default:) SCPI-conform behavior
The first command is always referenced to the root node. Subsequent
commands, however, are referenced to the same tree level as the
previous command in a message unit.
"Enhanced"
If subsequent commands are not found on the same tree level as the
previous command, the SCPI parser searches "backwards up the
tree" of defined SCPI commands. If the command is found at a higher
node, that command is used. Only if the command is not found any-
where, an error message is returned.
Remote command:
SYSTem:TWALking on page 993
Instrument Setup and Interface Configuration – SETUP Key

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals