S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_fetch_saudio.fm
FETCh:SAUDio
FETCh:SAUDio:DISTortion:MINimum?
FETCh:SAUDio:DISTortion:SDEViation?
FETCh:SAUDio:ICOunt?
FETCh:SAUDio:INTegrity?
Function This query returns a comma separated list of the minimum audio distortion measured at each
frequency point in the sweep.
For frequencies above 10 kHz, this measurement returns 9.91 E+37.
Query Range: 0 to 99.9%, 9.91 E+37
Resolution: 0.01%
Function This query returns a comma separated list of the standard deviation audio distortion measured at
each frequency point in the sweep. The number of frequency points is determined with the
SETup:SAUDio:FREQuency:POINts command.
For frequencies above 10 kHz, this measurement returns 9.91 E+37.
Query Range: 0 to 99.9%, 9.91 E+37
Resolution: 0.001%
Function This query returns the intermediate count of swept audio multi-measurements completed. The
total measurement count is the number of measurements specified in with the
multi-measurement count command, SETup:SAUDio:COUNt[:SNUMber] , times the number of
frequency points specified with the SETup:SAUDio:FREQuency:POINts . Therefore, when
completing a 5 point sweep with multi-measurements turned off, the measurement will return 5
not 1.
See also SETup:SAUDio:ICOunt:MAXimum?
Query Range: 1 to 999 x 60
Resolution: 1
Function This query returns the integrity indicator for the last swept audio measurement completed. Zero
indicates a normal measurement. See Integrity Indicator on page 222 for descriptions of
non-zero integrity indicators.
Query Range: See Integrity Indicator on page 222.
Resolution: 1