Compatible SCPI Commands
Supported SCPI Commands
DISPlay[:WINDow<1|2>]:TIME ON | OFF
Switches on-screen display of date and time on or off.
This command is accepted, but takes no action and reports no error.
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:Y[:SCALe] 10dB to 200dB
The command specifies the display range of the Y-axis (level axis) with logarithmic scaling
(DISP:TRAC:Y:SPAC LOG).
The query returns the current value of this setting.
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:Y[:SCALe]:MODE ABSolute | RELativ
Specifies the scale type of the Y-axis (absolute or relative).
Remote Command Notes FSE, FSP, FSU
Preset OFF
State Saved Saved in instrument state.
Backwards Compatibility SCPI :DISPlay:TEXT:STATe
Remote Command :DISPlay:WINDow[1]:TIME ON|OFF|1|0
:DISPlay:WINDow[1]:TIME?
Example DISP:WIND:TIME ON
DISP:WIND:TIME?
Remote Command Notes FSE, FSP, FSU
Preset OFF
State Saved Saved in instrument state.
Backwards Compatibility SCPI :DISPlay:TIME
Remote Command DISPlay:WINDow[1]:TRACe[1]|2|3:Y[:SCALe] <rel_ampl>
DISPlay:WINDow[1]:TRACe[1]|2|3:Y[:SCALe]?
Example DISP:WIND:TRAC:Y 20dB
DISP:WIND:TRAC:Y?
Remote Command Notes FSE, FSP, FSU
Preset 100dB
Min 10dB
Max 200dB
Backwards Compatibility SCPI :DISPlay[1]:TRACe[1]|2|3:Y[:SCALe]