Compatible SCPI Commands
Supported SCPI Commands
DISPlay[:WINDow<1|2>]:TIME ON | OFF
Switches on-screen display of date and time on or off.
This command is accepted, but takes no action and reports no error.
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:Y[:SCALe] 10dB to 200dB
The command specifies the display range of the Y-axis (level axis) with logarithmic scaling
(DISP:TRAC:Y:SPAC LOG).
The query returns the current value of this setting.
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:Y[:SCALe]:MODE ABSolute | RELativ
Specifies the scale type of the Y-axis (absolute or relative).
Remote Command Notes FSE, FSP, FSU
Preset OFF
State Saved Saved in instrument state.
Backwards Compatibility SCPI :DISPlay:TEXT:STATe
Remote Command :DISPlay:WINDow[1]:TIME ON|OFF|1|0
:DISPlay:WINDow[1]:TIME?
Example DISP:WIND:TIME ON
DISP:WIND:TIME?
Remote Command Notes FSE, FSP, FSU
Preset OFF
State Saved Saved in instrument state.
Backwards Compatibility SCPI :DISPlay:TIME
Remote Command DISPlay:WINDow[1]:TRACe[1]|2|3:Y[:SCALe] <rel_ampl>
DISPlay:WINDow[1]:TRACe[1]|2|3:Y[:SCALe]?
Example DISP:WIND:TRAC:Y 20dB
DISP:WIND:TRAC:Y?
Remote Command Notes FSE, FSP, FSU
Preset 100dB
Min 10dB
Max 200dB
Backwards Compatibility SCPI :DISPlay[1]:TRACe[1]|2|3:Y[:SCALe]
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals