Compatible SCPI Commands
Supported SCPI Commands
When this function is activated, the APD measurement is turned off.
The query returns the current state of this setting.
CALCulate:STATistics:CCDF:X<1â€Ļ3>? P0_01 | P0_1 | P1 | P10
Queries the level values for the probabilities 0.01%, 0.1%, 1% and 10%.
CALCulate:STATistics:NSAMples <100â€Ļ1E9>
The command specifies the number of measurement points to be acquired for the statistical measurement
functions.
The minimum value supported by this mode is 1000.
The query returns the current value of this setting.
Remote Command :CALCulate:STATistics:CCDF:STATe ON|OFF|1|0
:CALCulate:STATistics:CCDF:STATe?
Example CALC:STAT:CCDF:STAT ON
Remote Command Notes FSP, FSU
Preset OFF
State Saved Saved in instrument state.
Remote Command CALCulate:STATistics:CCDF:X[1]|2|3? P0_01|P0_1|P1|P10
Example CALC:STAT:CCDF:X? P10
Remote Command Notes FSU
Remote Command :CALCulate[1]:STATistics:NSAMples <int>
:CALCulate[1]:STATistics:NSAMples?
Example CALC:STAT:NSAM 100000
CALC:STAT:NSAM?
Remote Command Notes FSP, FSU
Preset 100000
State Saved Saved in instrument state.
Min 1000
Max 1E9
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals