Compatible SCPI Commands
Supported SCPI Commands
[SENSe<1|2>:]FREQuency:STARt 0 to fmax
Specifies the analyzer’s start frequency. This command is available only in the frequency domain (span
> 0).
[SENSe<1|2>:]FREQuency:STOP 0 to fmax
Specifies the analyzers stop frequency. This command is available only in the frequency domain (span >
0).
Example FREQ:SPAN:FULL
Remote Command Notes FSE, FSP, FSU
Backwards Compatibility SCPI :SENSe1:FREQuency:SPAN:FULL
Remote Command [:SENSe]:FREQuency:STARt <freq>
[:SENSe]:FREQuency:STARt?
Example FREQ:STAR 200MHz
FREQ:STAR?
Remote Command Notes FSE, FSP, FSU
Preset 0 Hz
State Saved Saved in instrument state.
Min 0Hz
Max Depends on the instrument maximum frequency
Remote Command [:SENSe]:FREQuency:STOP <freq>
[:SENSe]:FREQuency:STOP?
Example FREQ:STOP 220MHz
FREQ:STOP?
Remote Command Notes FSE, FSP, FSU
Preset 3GHZ
State Saved Saved in instrument state.
Min 0Hz
Max 27.0 GHz