Page #141 background image
140
Compatible SCPI Commands
Supported SCPI Commands
[SENSe<1|2>:]FREQuency:STARt 0 to fmax
Specifies the analyzer’s start frequency. This command is available only in the frequency domain (span
> 0).
[SENSe<1|2>:]FREQuency:STOP 0 to fmax
Specifies the analyzer’s stop frequency. This command is available only in the frequency domain (span >
0).
Example FREQ:SPAN:FULL
Remote Command Notes FSE, FSP, FSU
Backwards Compatibility SCPI :SENSe1:FREQuency:SPAN:FULL
Remote Command [:SENSe]:FREQuency:STARt <freq>
[:SENSe]:FREQuency:STARt?
Example FREQ:STAR 200MHz
FREQ:STAR?
Remote Command Notes FSE, FSP, FSU
Preset 0 Hz
State Saved Saved in instrument state.
Min 0Hz
Max Depends on the instrument maximum frequency
Remote Command [:SENSe]:FREQuency:STOP <freq>
[:SENSe]:FREQuency:STOP?
Example FREQ:STOP 220MHz
FREQ:STOP?
Remote Command Notes FSE, FSP, FSU
Preset 3GHZ
State Saved Saved in instrument state.
Min 0Hz
Max 27.0 GHz
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals