33
Front Panel Operations
Mode Setup
Alternatively, you can select the desired response using the following SCPI command.
Compatibility Model
Sets the response to *IDN command as the emulated instrument, like
Rohde&Schwarz,FSP-26,123456/789,04.00.
System IDN Response
Sets the response to *IDN command as the analyzer system, like “N9020A”.
User IDN Response
Allows you to set or get the user-defined mode IDN response.
If you don’t enter anything, the system IDN response will be used instead.
Key Path: Mode Setup
Remote Command: [:SENSe]:RLC:IDN:TYPE SYSTem|COMPatibility|USER
[:SENSe]:RLC:IDN:TYPE?
Example: SENS:RLC:IDN:TYPE SYSTem
RLC:IDN:TYPE?
State Saved:
Saved in instrument state.
Range: SYSTem | COMPatibility | USER
Initial S/W Revision: A.06.00 or later
Key Path:
Mode Setup, Mode IDN Response
Example: RLC:IDN:TYPE COMP
Initial S/W Revision: A.06.00 or later
Key Path:
Mode Setup, Mode IDN Response
Example: RLC:IDN:TYPE SYST
Initial S/W Revision: A.06.00 or later
Key Path:
Mode Setup
Remote Command: [:SENSe]:RLC:IDN:USER <string>
[:SENSe]:RLC:IDN:USER?
Example: RLC:IDN:USER “abc”
RLC:IDN:USER?
State Saved:
Saved in instrument state.
Initial S/W Revision: A.06.00 or later