82
Compatible SCPI Commands
Supported SCPI Commands
CALCulate<1|2>:LIMit<1...6>:LOWer:SPACing LINear | LOGarithmic
The command switches between linear or logarithmic interpolation for the lower limit line.
The query returns the current state of this setting.
CALCulate<1|2>:LIMit<1...6>:LOWer:THReshold <numeric_value>
The command defines an absolute threshold value for limit lines with relative Y-axis scaling.
The absolute threshold value is used for the limit check when it exceeds the relative limit value.
This mode supports only dBm units.
The query returns the current value of this setting.
CALCulate:LIMit:UPPer Subsystem
The CALCulate:LIMit:UPPer subsystem controls the upper limit line.
Example CALC:LIM:LOW:SHIF 1
CALC:LIM:LOW:SHIF?
Remote Command Notes FSE, FSP, FSU
Preset 0
Remote Command :CALCulate[1]:LIMit[1]|2|3|4|5|6:LOWer:SPACing
LINear|LOGarithmic
:CALCulate[1]:LIMit[1]|2|3|4|5|6:LOWer:SPACing?
Example CALC:LIM:LOW:SPAC LIN
CALC:LIM:LOW:SPAC?
Remote Command Notes FSE, FSP, FSU
Preset LIN
State Saved
Saved in instrument state.
Range LINear | LOGarithmic
Remote Command CALCulate[1]:LIMit[1]|2|3|4|5|6:LOWer:THReshold <ampl>
CALCulate[1]:LIMit[1]|2|3|4|5|6:LOWer:THReshold?
Example CALC:LIM:LOW:THR 1DBM
CALC:LIM:LOW:THR?
Remote Command Notes FSP, FSU
Preset â200dBm