143
Compatible SCPI Commands
Supported SCPI Commands
adjacent-channel power measurements.
[SENSe<1|2>:]POWer:ACHannel:ACPairs 0 | 1 | 2 | 3
Specifies the number of adjacent channels (upper and lower channel in pairs).The number 0 represents a
pure channel power measurement.
[SENSe<1|2>:]POWer:ACHannel:BANDwidth|BWIDth[:CHANnel] 100 Hz to 1000 MHz
Specifies the radio communication system’s channel bandwidth.
Remote Command [:SENSe]:POWer:ACHannel:TXCHannel:COUNt <int>
[:SENSe]:POWer:ACHannel:TXCHannel:COUNt?
Example POW:ACH:TXCH:COUN 1
POW:ACH:TXCH:COUN?
Remote Command Notes FSP, FSU
Preset 4
State Saved Saved in instrument state.
Min 1
Max 4
Remote Command [:SENSe]:POWer:ACHannel:ACPairs
[:SENSe]:POWer:ACHannel:ACPairs?
Example POW:ACH:ACP 3
Remote Command Notes FSE, FSP, FSU
Preset 0
Range 0|1|2|3
Remote Command [:SENSe]:POWer:ACHannel:BANDwidth|BWIDth[:CHANnel]
<freq>
[:SENSe]:POWer:ACHannel:BANDwidth|BWIDth[:CHANnel]?
Example POW:ACH:BWID 30kHz
Remote Command Notes FSE, FSP, FSU
Preset 14KHz
State Saved Saved in instrument state.
Min 100Hz
Max 1000MHz
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals