152
Compatible SCPI Commands
Supported SCPI Commands
STATus Subsystem
The STATus subsystem provides the status reporting system commands. The status registers are not
influenced by *RST.
STATus:OPERation[:EVENt]?
Queries the contents of the EVENt section in the STATus:OPERation register. After the readout, the
contents of the EVENt section are deleted.
STATus:OPERation:CONDition?
Queries the CONDition section in the STATus:OPERation register. The current hardware status is
reflected in the value returned. The contents of the CONDition section are not deleted after the readout.
STATus:OPERation:ENABle
Specifies the bits of the ENABle section in the STATus:OPERation register.
This mode supports the maximum value of 32767, not 65535 as in FSx.
State Saved Saved in instrument state.
Min 1
Max 8001
Remote Command :STATus:OPERation[:EVENt]?
Example STAT:OPER?
Remote Command Notes FSE, FSP, FSU
Preset 0
Remote Command :STATus:OPERation:CONDition
Example STAT:OPER:COND?
Remote Command Notes FSE, FSP, FSU
Preset 0
Remote Command :STATus:OPERation:ENABle <integer>
:STATus:OPERation:ENABle?
Example STAT:OPER:ENAB 1
Remote Command Notes FSE, FSP, FSU
Preset 0
Min 0
Max 32767
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals