136
Compatible SCPI Commands
Supported SCPI Commands
[SENSe<1|2>:]CORRection:TRANsducer[:STATe] ON | OFF
Switches the selected transducer factor on or off.
Currently, this command support 6 correction settings.
[SENSe<1|2>:]CORRection:TRANsducer:DELete
Deletes the selected transducer factor.
Currently, this command support 6 correction settings.
[SENSe<1|2>:]CORRection:TRANsducer:VIEW ON | OFF
Switches on the display of the active transducer factor or set.
Currently, this command support 6 correction settings.
Example CORR:TRAN:DATA 1,2,3,4,5,6,7,8,9,0
CORR:TRAN:DATA?
Notes All parameters are sent as dimensionless. Currently, this command support 6
correction settings.
FSU, FSP, FSE
State Saved Saved in instrument state.
Remote Command [:SENSe]:CORRection:TRANsducer:STATe ON|OFF|1|0
[:SENSe]:CORRection:TRANsducer:STATe?
Example CORR:TRAN:STAT OFF
CORR:TRAN:STAT?
Notes FSU, FSP, FSE
Preset OFF
State Saved Saved in instrument state.
Remote Command [:SENSe]:CORRection:TRANsducer:DELete
Example CORR:TRAN:DEL
Notes FSU, FSP, FSE
State Saved Saved in instrument state.
Remote Command [:SENSe]:CORRection:TRANsducer:VIEW ON|OFF|1|0
[:SENSe]:CORRection:TRANsducer:VIEW?
Example CORR:TRAN:VIEW OFF
CORR:TRAN:VIEW?