64
Compatible SCPI Commands
Supported SCPI Commands
CALCulate<1|2>:DELTamarker<1...4>:LINK ON | OFF
The command switches the link between delta marker 1 and marker 1 on or off.
If the link is âonâ, and the X-axis value of marker 1 is changed, then delta marker 1 will change by the
same amount.
The numeric suffix of DELTamarker may only be 1, or be omitted, because this functionality is only
available for marker 1 and delta marker 1.
The query returns the current state of this setting.
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed[:STATe] ON | OFF
The command switches between relative measurement, and measurement relative to a fixed reference.
If required, marker 1 is activated and a peak search is performed.
If marker 1 is activated, its position becomes the reference point for the measurement. The reference
point can then be changed, using commands CALCulate:DELTamarker:FUNCtion:FIXed:RPOint:X and
:RPOint:Y, independently of the position of marker 1 and of any trace.
The query returns the current state of this setting.
Example CALC:DELT:MIN:RIGH
Remote Command Notes FSE, FSP, FSU
Remote Command :CALCulate[1]:DELTamarker[1]|2|3|4:LINK ON|OFF|1|0
:CALCulate[1]:DELTamarker[1]|2|3|4:LINK?
Example CALC:DELT:LINK ON
Remote Command Notes FSU
Preset OFF
State Saved Saved in instrument state.
Remote Command :CALCulate[1]:DELTamarker[1]|2|3|4:FUNCtion:FIXed[:STAT
e] ON|OFF|1|0
:CALCulate[1]:DELTamarker[1]|2|3|4:FUNCtion:FIXed[:STAT
e]?
Example CALC:DELT:FUNC:FIX ON
Remote Command Notes FSE, FSP, FSU
Preset OFF
State Saved Saved in instrument state.