163
Compatible SCPI Commands
Supported SCPI Commands
This mode supports the level range:–200 to 100 dBm.
TRIGger<1|2>[:SEQuence]:LEVel:VIDeo 0 to 100PCT
Sets the video trigger level.
TRIGger<1|2>[:SEQuence]:HOLDoff
Specifies the trigger delay period.
Remote Command :TRIGger[1][:SEQuence]:LEVel:RFPower <real>
:TRIGger[1][:SEQuence]:LEVel:RFPower?
Example TRIG:LEV:RFP –20DBM
Remote Command Notes FSP
Preset –20dBm
State Saved Saved in instrument state.
Min –200dBm
Max 100dBm
Remote Command TRIGger[1][:SEQuence]:LEVel:VIDeo <real>
TRIGger[1][:SEQuence]:LEVel:VIDeo?
Example TRIG:LEV:VID 40
TRIG:LEV:VID?
Remote Command Notes FSE, FSP, FSU
Preset 50
State Saved Saved in instrument state.
Min 0
Max 100
Remote Command :TRIGger[1][:SEQuence]:HOLDoff <time>
:TRIGger[1][:SEQuence]:HOLDoff?
Example TRIG:HOLD 100ms
Remote Command Notes FSE, FSP, FSU
Preset 0s
State Saved Saved in instrument state.
Min 0s
Max 0.5s
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals