133
Compatible SCPI Commands
Supported SCPI Commands
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:AUTO ON | OFF
Automatically couples the instrument's video bandwidth to its resolution bandwidth or cancels the
coupling.
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:RATio 0.01 to 1000
Specifies the video bandwidth (Hz) / resolution bandwidth (Hz) ratio.
SENSe:CORRection Subsystem
The SENSe:CORRection subsystem controls calibration andnormalization.
[SENSe<1|2>:]CORRection:EGAin:INPut[:MAGNitude] –200...200dB
Informs the analyzer of an external gain.
Min 1 Hz
Max 50 MHz
Backwards Compatibility SCPI :SENSe1:BANDwidth|BWIDth[:RESolution]:VIDeo
Remote Command [:SENSe]:BANDwidth|BWIDth:VIDeo:AUTO
ON|OFF|1|0[:SENSe]:BANDwidth|BWIDth:VIDeo:AUTO?
Example BAND:VID:AUTO OFF
Remote Command Notes FSE, FSP, FSU
Preset ON
State Saved Saved in instrument state.
Backwards Compatibility SCPI :SENSe1:BANDwidth|BWIDth[:RESolution]:VIDeo:AUTO
Remote Command [:SENSe]:BANDwidth|BWIDth:VIDeo:RATio <real>
[:SENSe]:BANDwidth|BWIDth:VIDeo:RATio?
Example BAND:VID:RAT 2
BAND:VID:RAT?
Remote Command Notes FSE, FSP, FSU
Preset 3
State Saved Saved in instrument state.
Min 0.00001
Max 3000000
Backwards Compatibility SCPI :SENSe1:BANDwidth|BWIDth[:RESolution]:VIDeo:RATio
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals