138
Compatible SCPI Commands
Supported SCPI Commands
either be set via the center frequency and span, or via the start and stop frequency.
[SENSe<1|2>:]FREQuency:CENTer 0 to fmax
Specifies the analyzerâs center frequency or the measuring frequency for span = 0.
[SENSe<1|2>:]FREQuency:CENTer:STEP 0 to fmax
Specifies the center frequencyâs step width.
[SENSe<1|2>:]FREQuency:CENTer:STEP:LINK SPAN | RBW | OFF
Couples the center frequencyâs step width to the span (span >0) or to the resolution bandwidth (span =
0), or cancels the coupling.
Remote Command [:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer?
Example FREQ:CENT 50MHz
FREQ:CENT?
Remote Command Notes FSE, FSP, FSU
Preset fmax /2 with fmax = maximum frequency
State Saved Saved in instrument state.
Min Depends on instrument minimum frequency
Max Depends on instrument maximum frequency
Backwards Compatibility SCPI :SENSe1:FREQuency:CENTer
Remote Command [:SENSe]:FREQuency:CENTer:STEP <freq>
[:SENSe]:FREQuency:CENTer:STEP?
Example FREQ:CENT:STEP 120MHz
Remote Command Notes FSE, FSP, FSU
Preset 1MHz
State Saved Saved in instrument state.
Min 1 Hz
Max 27.0 GHz
Backwards Compatibility SCPI :SENSe1:FREQuency:CENTer:STEP
Remote Command [:SENSe]:FREQuency:CENTer:STEP:LINK SPAN | RBW | OFF
[:SENSe]:FREQuency:CENTer:STEP:LINK?