Compatible SCPI Commands
Supported SCPI Commands
CALCulate<1|2>:THReshold MINimum to MAXimum (depending on current unit)
The command specifies the threshold value for the maximum/minimum search of markers with marker
search functions MAX PEAK, NEXT PEAK, etc.
The associated display line is automatically turned on.
The query returns the current value of this setting.
CALCulate<1|2>:THReshold:STATe ON | OFF
The command turns the threshold line on or off.
The unit depends on the setting performed by CALC:UNIT.
The query returns the current state of this setting.
CALCulate<1|2>:FLINe<1|2> 0...fmax
Specifies the position of the frequency lines.
This command is accepted, but takes no action and reports no error.
CALCulate<1|2>:FLINe<1|2>:STATe ON | OFF
Switches the frequency line on or off.
Remote Command :CALCulate[1]:THReshold <ampl>
:CALCulate[1]:THReshold?
Example CALC:THR â82DBM
Remote Command Notes FSE, FSP, FSU
State Saved Saved in instrument state.
Remote Command :CALCulate[1]:THReshold:STATe ON|OFF|1|0
:CALCulate[1]:THReshold:STATe?
Example CALC:THR:STAT ON
Remote Command Notes FSE, FSP, FSU
Preset OFF
State Saved Saved in instrument state.
Remote Command :CALCulate[1]:FLINe[1]|2 <freq>
:CALCulate[1]:FLINe[1]|2?
Example CALC:FLIN 120MHz
Remote Command Notes FSE, FSP, FSU
State Saved Saved in instrument state.