Compatible SCPI Commands
Supported SCPI Commands
SENSe:DETector Subsystem
The SENSe:DETector subsystem controls measurement data acquisition via selection of the detector for
the corresponding trace.
[SENSe<1|2>:]DETector<1..3>[:FUNCtion] APEak | NEGative | POSitive | SAMPle | RMS
|AVERage | QPEak
Switches on the data acquisition detector in the selected trace.
[SENSe<1|2>:]DETector<1 to 3>[:FUNCtion]:AUTO ON | OFF
Either couples the detector to the current trace setting or turns coupling off.
SENSe:FREQuency Subsystem
The SENSe:FREQuency subsystem specifies the active display’s frequency axis. The frequency axis can
Notes FSU, FSP, FSE
Preset OFF
State Saved Saved in instrument state.
Remote Command [:SENSe]:DETector[1]|2|3[:FUNCtion]
APEak|NEGative|POSitive|SAMPle|RMS|AVERage|QPEak
[:SENSe]:DETector[1]|2|3[:FUNCtion]?
Example DET POS
Remote Command Notes
Remote Command Notes FSE, FSP, FSU
Preset APEak
State Saved Saved in instrument state.
Range APEak | NEGative | POSitive | SAMPle | RMS | AVERage | QPEak
Backwards Compatibility SCPI :SENSe1:DETector[1]|2|3[:FUNCtion]
Remote Command [:SENSe]:DETector[1]|2|3[:FUNCtion]:AUTO ON|OFF|1|0
[:SENSe]:DETector[1]|2|3[:FUNCtion]:AUTO?
Example DET:AUTO OFF
Remote Command Notes FSE, FSP, FSU
Preset ON
State Saved Saved in instrument state.
Backwards Compatibility SCPI :SENSe1:DETector[1]|2|3[:FUNCtion]:AUTO