EasyManua.ls Logo

Agilent Technologies EXA N9010A - TRIGger Subsystem

Agilent Technologies EXA N9010A
175 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
162
Compatible SCPI Commands
Supported SCPI Commands
TRIGger Subsystem
This subsystem includes commands that control and synchronize the start of a sweep.
An external trigger signal can be supplied via the connector on the analyzer’s rear panel.
TRIGger<1|2>[:SEQuence]:SOURce
Selects the trigger source.
This mode supports only parameter options: IMMediate, EXTernal, VIDeo, RFPower.
It does not support parameter options: IFPower, TV, AF, FM, AM and PM.
TRIGger<1|2>[:SEQuence]:LEVel:IFPower –30 to –10DBM
This command sets the level of the IF power trigger source.
This command is accepted, but takes no action and reports no error.
TRIGger<1|2>[:SEQuence]:LEVel:RFPower
Sets the RF power trigger level.
Preset OFF
State Saved Saved in instrument state.
Remote Command :TRIGger[1][:SEQuence]:SOURce
IMMediate|EXTernal|VIDeo|IFPower|RFPower|TV|AF|FM|AM|PM
:TRIGger[1][:SEQuence]:SOURce?
Example TRIG:SOUR EXT
Remote Command Notes
Remote Command Notes FSE, FSP, FSU
Preset IMMediate
Range IMMediate|EXTernal|VIDeo|IFPower|RFPower|TV|AF|FM|AM|PM
Remote Command :TRIGger[1][:SEQuence]:LEVel:IFPower <ampl>
:TRIGger[1][:SEQuence]:LEVel:IFPower?
Example TRIG:LEV:IFP –20DBM
Remote Command Notes FSP, FSU
Preset –20dBm
State Saved Saved in instrument state.
Min 30dBm
Max –70dBm