Compatible SCPI Commands
Supported SCPI Commands
[SENSe<1|2>:]FREQuency:MODE CW | FIXed | SWEep
This command switches between time domain (CW | FIXed) and frequency domain (SWEep).
[SENSe<1|2>:]FREQuency:OFFSet
Specifies the instrument’s frequency offset.
SENSe:POWer Subsystem
The SENSe:POWer subsystem controls the instrument's channel and adjacent channel power
measurements.
[SENSe<1|2>:]POWer:ACHannel:SPACing:CHANnel 100 Hz to 2000 MHz
Specifies the channel spacing for the carrier signals. This command is available only for measurements
in the frequency domain (span > 0).
Remote Command [:SENSe]:FREQuency:MODE CW|FIXed|SWEep
[:SENSe]:FREQuency:MODE?
Example FREQ:MODE CW
Remote Command Notes FSE, FSP, FSU
Preset SWEep
Range CW|FIXed|SWEep
Backwards Compatibility SCPI :SENSe1:FREQuency:MODE
Remote Command [:SENSe]:FREQuency:OFFSet <freq>
[:SENSe]:FREQuency:OFFSet?
Example FREQ:OFFS 10MHz
Remote Command Notes FSE, FSP, FSU
Preset 0 Hz
State Saved Saved in instrument state.
Min –500 GHz
Max 500 GHz
Backwards Compatibility SCPI :SENSe1:FREQuency:OFFSet
Remote Command [:SENSe]:POWer:ACHannel:SPACing:CHANnel 1000
[:SENSe]:POWer:ACHannel:SPACing:CHANnel
Example POW:ACH:SPAC:CHAN 2000
POW:ACH:SPAC:CHAN?