EasyManua.ls Logo

Agilent Technologies EXA N9010A - Status:operation:ptransition; Status:operation:ntransition; Status:preset

Agilent Technologies EXA N9010A
175 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
153
Compatible SCPI Commands
Supported SCPI Commands
STATus:OPERation:PTRansition
Specifies the edge detectors of all bits in the STATus:OPERation register from 0 to 1 for the transitions
of the CONDition bit.
This mode supports the maximum value of 32767, not 65535 as FSx.
STATus:OPERation:NTRansition
Specifies the edge detectors of all bits in the STATus:OPERation register from 1 to 0 for the transitions
of the CONDition bit.
This mode supports the maximum value of 32767, not 65535 as in FSx.
STATus:PRESet
Resets the edge detectors and ENABle parts of all registers to a defined value.
Remote Command :STATus:OPERation:PTRansition <integer>
:STATus:OPERation:PTRansition?
Example STAT:OPER:PTR 1
Remote Command Notes FSE, FSP, FSU
Preset 32767
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All
Remote Command :STATus:OPERation:NTRansition <integer>
:STATus:OPERation:NTRansition?
Example STAT:OPER:NTR 1
Remote Command Notes FSE, FSP, FSU
Preset 0
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Remote Command :STATus:PRESet
Example STAT:PRES