EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>EXA N9010A

Agilent Technologies EXA N9010A User Manual

Agilent Technologies EXA N9010A
175 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #73 background imageLoading...
Page #73 background image
72
Compatible SCPI Commands
Supported SCPI Commands
CALCulate<1|2>:LIMit<1...6>:ESPectrum:RESTore
Restores the standard limit lines for the spectrum emission mask measurement. Any modifications that
were made to the standard limit lines are lost, and the default setting is restored.
This command is accepted, but takes no action and reports no error.
CALCulate:LIMit:ACPower Subsystem
The CALCulate:LIMit:ACPower subsystem checks the limit for adjacent channel power (ACP)
measurement.
CALCulate<1|2>:LIMit<1â€Ļ6>:ACPower[:STATe]
Switches on and off the limit check for ACP measurements in the selected measurement window. In
addition, you should use the commands CALCulate:LIMit:ACPower:ACHannel:STATe or
CALCulate:LIMit:ACPower:ALTernate:STATe to specify whether the limit check is to be performed for
the upper/lower adjacent channel or for the alternate adjacent channels.
The numeric suffix is irrelevant for this command.
CALCulate<1|2>:LIMit<1â€Ļ6>:ACPower:ACHannel[:RELative]
Specifies the relative limit of the upper/lower adjacent channel for ACP measurements in the selected
measurement window. The reference value for the relative limit value is the measured channel power.
The numeric suffix is irrelevant for this command.
Remote Command :CALCulate[1]:LIMit[1]|2|3|4|5|6:ESPectrum:RESTore
Example CALC:LIM:ESP:REST
Remote Command Notes FSP
Remote Command :CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower[:STATe]
ON|OFF|1|0
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower[:STATe]?
Example CALC:LIM:ACP ON
CALC:LIM:ACP?
Notes FSU, FSP, FSE
Preset OFF
State Saved Saved in instrument state.
Remote Command :CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower:ACHannel[:RELa
tive] <real>,<real>
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower:ACHannel[:RELa
tive]?
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals