Supported SCPI Commands
The query returns the current value of this setting.
INPut<1|2>:ATTenuation:AUTO
The command switches between coupling of the input attenuation to the reference level (state ON), and
manual input attenuation (state OFF).
The query returns the current state of this setting.
INPut<1|2>:EATT
The command specifies the setting of the electronic input attenuator.
This mode supports a maximum value of 24 dB.
The query returns the current value of this setting.
Remote Command :INPut[1]:ATTenuation <rel_ampl>
:INPut[1]:ATTenuation?
Example INP:ATT 40dB
Remote Command Notes FSE, FSP, FSU
Preset 6dB
State Saved Saved in instrument state.
Min 0
Max 70
Remote Command :INPut[1]:ATTenuation:AUTO ON|OFF|1|0
:INPut[1]:ATTenuation:AUTO?
Example INP:ATT:AUTO ON
Remote Command Notes FSP, FSU
Preset ON
State Saved Saved in instrument state.
Remote Command :INPut[1]:EATT <rel_ampl>
:INPut[1]:EATT?
Example INP:EATT 15dB
Remote Command Notes FSP, FSU
Preset 0
State Saved Saved in instrument state.
Min 0
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals