93
Compatible SCPI Commands
Supported SCPI Commands
The query returns the current value of this setting.
CALCulate:MARKer:FUNCtion Subsystem
CALCulate<1|2>:MARKer<1…4>:FUNCtion:CENTer
Sets the center frequency of the selected measurement window equal to the frequency of the indicated
marker.
CALCulate<1|2>:MARKer<1…4>:FUNCtion:CSTep
Sets the step width of the center frequency in the selected measurement window to the X value of the
current marker.
CALCulate<1|2>:MARKer<1…4>:FUNCtion:FPEaks
Searches the selected trace for the indicated number of maxima.
Remote Command :CALCulate[1]:MARKer[1]|2|3|4:PEXCursion <rel_ampl>
:CALCulate[1]:MARKer[1]|2|3|4:PEXCursion?
Example CALC:MARK:PEXC 10dB
Remote Command Notes FSE, FSP, FSU
Preset 6dB
State Saved Saved in instrument state.
Remote Command :CALCulate[1]:MARKer[1]|2|3|4:FUNCtion:CENTer
Example CALC:MARK:FUNC:CENT
Notes FSE, FSP, FSU
Remote Command :CALCulate[1]:MARKer[1]|2|3|4:FUNCtion:CSTep
Example CALC:MARK:FUNC:CST
Notes FSE, FSP, FSU
Remote Command :CALCulate[1]:MARKer[1]|2|3|4:FUNCtion:FPEaks[:IMMediat
e] <int>
:CALCulate[1]:MARKer[1]|2|3|4:FUNCtion:FPEaks[:IMMediat
e]?
Example CALC:MARK:FUNC:FPE 5
Notes FSE, FSP, FSU
Preset 0