Compatible SCPI Commands
Supported SCPI Commands
INITiate Subsystem
The INITiate subsystem controls the init-measurement function.
INITiate<1|2>:CONTinuous ON | OFF
The command switches the trigger system between the continuously-initiated mode (Continuous) or
single measurement mode (Single).
The query returns the current state of this setting.
INITiate<1|2>:CONMeas
In Single sweep mode, this command resumes a stopped measurement from the current position.
INITiate<1|2>[:IMMediate]
Initiates a new sweep.
INPut Subsystem
The INPut subsystem controls the input characteristics of the instruments RF inputs.
INPut<1|2>:ATTenuation
The command specifies the input attenuator setting.
With the Electronic Attenuator Option, this mode supports 2 dB steps.
Remote Command :INITiate[1]:CONTinuous ON|OFF|1|0
:INITiate[1]:CONTinuous?
Example :INIT:CONT 0
:INIT:CONT 1
Remote Command Notes FSE, FSP, FSU
Preset ON
State Saved Saved in instrument state.
Remote Command :INITiate[1]:CONMeas
Example INIT:CONM
Remote Command Notes FSE, FSP, FSU
Remote Command :INITiate[1][:IMMediate]
Example INIT
Remote Command Notes FSE, FSP, FSU