11 8
Compatible SCPI Commands
Supported SCPI Commands
INITiate Subsystem
The INITiate subsystem controls the init-measurement function.
INITiate<1|2>:CONTinuous ON | OFF
The command switches the trigger system between the continuously-initiated mode (Continuous) or
single measurement mode (Single).
The query returns the current state of this setting.
INITiate<1|2>:CONMeas
In Single sweep mode, this command resumes a stopped measurement from the current position.
INITiate<1|2>[:IMMediate]
Initiates a new sweep.
INPut Subsystem
The INPut subsystem controls the input characteristics of the instrument’s RF inputs.
INPut<1|2>:ATTenuation
The command specifies the input attenuator setting.
With the Electronic Attenuator Option, this mode supports 2 dB steps.
Remote Command :INITiate[1]:CONTinuous ON|OFF|1|0
:INITiate[1]:CONTinuous?
Example :INIT:CONT 0
:INIT:CONT 1
Remote Command Notes FSE, FSP, FSU
Preset ON
State Saved Saved in instrument state.
Remote Command :INITiate[1]:CONMeas
Example INIT:CONM
Remote Command Notes FSE, FSP, FSU
Remote Command :INITiate[1][:IMMediate]
Example INIT
Remote Command Notes FSE, FSP, FSU
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals