Compatible SCPI Commands
Supported SCPI Commands
INPut<1|2>:EATT:AUTO
The command switches between coupling of the electronic input attenuation to the reference level (state
ON), and manual input attenuation (state OFF).
The query returns the current state of this setting.
INPut<1|2>:EATT:STATe
The command switches electronic input attenuation on or off.
The query returns the current state of this setting.
INPut<1|2>:IMPedance
The command sets the instrumentâs nominal input impedance.
The query returns the current state of this setting.
Max 24
Remote Command :INPut[1]:EATT:AUTO ON|OFF|1|0
:INPut[1]:EATT:AUTO?
Example INP:EATT:AUTO ON
Remote Command Notes FSP, FSU
Preset ON
State Saved Saved in instrument state.
Remote Command :INPut[1]:EATT:STATe ON|OFF|1|0
:INPut[1]:EATT:STATe?
Example INP:EATT:STAT ON
Remote Command Notes FSP, FSU
Preset OFF
State Saved Saved in instrument state.
Remote Command :INPut[1]:IMPedance 50|75
:INPut[1]:IMPedance?
Example INP:IMP 75
INP:IMP?
Remote Command Notes FSE, FSP, FSU
State Saved Saved in instrument state.
Range 50 | 75