Compatible SCPI Commands
Supported SCPI Commands
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
<numeric_value>
Sets the reference point for all delta markers when you are making measurements using a fixed reference
point (enabled by the command CALC:DELT:FUNC:FIX:STAT ON). You can specify a value using the
<numeric_value> parameter, or specify the peak of the selected trace [:PEAK].
For phase-noise measurements (selected via CALCulate:DELTamarker:FUNCtion:PNOise:STATe ON),
the command specifies a new reference point for delta marker 2.
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:Y <numeric_value>
Sets or queries a Y-axis reference point for all delta markers when you are making measurements using a
fixed reference point (enabled by CALCulate:DELTamarker:FUNCtion:FIXed:STATe ON).
For phase-noise measurements (selected via CALCulate:DELTamarker:FUNCtion:PNOise:STATe ON),
the command specifies a new reference point level for delta marker 2.
CALCulate<1|2>:DELTamarker<1...4>:FUNCtion:FIXed:RPOint:Y:OFFSet <numeric_value>
Sets or queries the additional offset for all delta markers when you are making measurements using a
fixed reference value (enabled by CALCulate:DELTamarker:FUNCtion:FIXed:STATe ON).
For phase-noise measurements (selected via CALCulate:DELTamarker:FUNCtion:PNOise:STATe ON),
specifies an additional offset that is included in the display of delta marker 2.
Remote Command :CALCulate[1]:DELTamarker[1]|2|3|4:FUNCtion:FIXed:RPOin
t:MAXimum[:PEAK]
Example CALC:DELT:FUNC:FIX:RPO:MAX
Remote Command Notes FSP, FSU
Remote Command :CALCulate[1]:DELTamarker[1]|2|3|4:FUNCtion:FIXed:RPOin
t:Y <amplitude>
:CALCulate[1]:DELTamarker[1]|2|3|4:FUNCtion:FIXed:RPOin
t:Y?
Example CALC:DELT2:FUNC:FIX:RPO:Y –10DBM
CALC:DELT2:FUNC:FIX:RPO:Y?
Remote Command Notes FSE, FSP, FSU
Remote Command :CALCulate[1]:DELTamarker[1]|2|3|4:FUNCtion:FIXed:RPOin
t:Y:OFFSet <rel_ampl>
:CALCulate[1]:DELTamarker[1]|2|3|4:FUNCtion:FIXed:RPOin
t:Y:OFFSet?
Example CALC:DELT2:FUNC:FIX:RPO:Y:OFFS 10DB
CALC:DELT2:FUNC:FIX:RPO:Y:OFFS?
Remote Command Notes FSE, FSP, FSU
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals