Practical CP/MAS Spectroscopy on Spin 1/2 Nuclei
User Manual Version 002 BRUKER BIOSPIN 103 (327)
- subtract 2 dB from pl1 (more is not required since the transmitter will usually
put out 50% more power at
15
N frequency)
- set p15 = 3 ms
- acquire 4-8 scans
- optimize HH condition, acquire reference spectrum with aq=25-35 ms
2.
29
Si on DSS
- load α-glycine
13
C reference spectrum
- set observe nucleus to
29
Si in edasp
- add 2 dB to sp0
- acquire 4 scans with aq = 35 ms
- optimize HH condition, acquire reference spectrum
3.
31
P on ADP (ammonium dihydrogen phosphate NH
4
H
2
PO
4
)
- load α-glycine
13
C reference spectrum
- set observe nucleus to
31
P in edasp
- add 6 dB to pl1
- optimize HH condition, acquire 2 scans, reduce rg appropriately.