280 (327) BRUKER BIOSPIN User Manual Version 002
CRAMPS 1D
l11=anavpt 4 2, 4, 8, 16 or 32.
o1p 10 or -1 To be optimized.
swh 1e
6
/2*(2*p9+10*p5)*0.6 To be corrected for proper scaling.
rg 16-64
td 512 Up to 1024.
si 4k
digmod analog
MASR 12-15 kHz Depending on cycle time.
Table 22.3. DUMBO, Analog Mode
Parameter Value Comment
pulprog dumboa Runs on AV 3 instruments only.
pl12
sp1
for 100 kHz RF field
up to 130 kHz To be optimized during setup.
spnam1 dumbo1_64 Set by xau dumbo.
p1 2.5 µsec For 100 kHz RF field.
p8 1.2µsec
p14 0.7 µsec To be optimized.
cnst25 140 To be optimized.
p9 4 – 2.6 µsec To be optimized.
p10 32 µsec or 24 µsec Set by xau dumbo.
d1 4s For α-glycine.
l11=anavpt 4 2, 4, 8, 16 or 32.
o1p 5 To be optimized.
swh 1e
6
/2*(2*p9+p10)*0.5 To be corrected for proper scaling.
rg 16-64
td 512 Up to 1024.
si 4k
digmod analog
MASR 10-12 kHz Depending on cycle time.
Table 22.2. PMLG Analog Mode

Table of Contents