Section 7.  Installation 
 
 
potentials at the two instruments.    For this reason, a differential measurement 
should be made on the analog output from the external signal conditioner. 
 
7.3.4  Ground Looping in Ionic Measurements 
When measuring soil-moisture with a resistance block, or water conductivity with 
a resistance cell, the potential exists for a ground loop error.    In the case of an 
ionic soil matric potential (soil moisture) sensor, a ground loop arises because soil 
and water provide an alternate path for the excitation to return to CR800 ground.   
This example is modeled in the diagram Model of a Ground Loop with a Resistive 
Sensor
 (p. 102). With R
g
 in the resistor network, the signal measured from the sensor 
is described by the following equation: 
 
where 
V
x
 is the excitation voltage 
R
f
 is a fixed resistor 
R
s
 is the sensor resistance 
R
g
 is the resistance between the excited electrode and CR800 earth ground. 
R
x
R
f
/R
g
 is the source of error due to the ground loop. When R
g
 is large, the error 
is negligible.    Note that the geometry of the electrodes has a great effect on the 
magnitude of this error.    The Delmhorst gypsum block used in the Campbell 
Scientific 227 probe has two concentric cylindrical electrodes.    The center 
electrode is used for excitation; because it is encircled by the ground electrode, the 
path for a ground loop through the soil is greatly reduced.    Moisture blocks which 
consist of two parallel plate electrodes are particularly susceptible to ground loop 
problems.    Similar considerations apply to the geometry of the electrodes in 
water conductivity sensors. 
The ground electrode of the conductivity or soil moisture probe and the CR800 
earth ground form a galvanic cell, with the water/soil solution acting as the 
electrolyte.    If current is allowed to flow, the resulting oxidation or reduction will 
soon damage the electrode, just as if dc excitation was used to make the 
measurement.    Campbell Scientific resistive soil probes and conductivity probes 
are built with series capacitors to block this dc current.    In addition to preventing 
sensor deterioration, the capacitors block any dc component from affecting the 
measurement.