Section 5.    Overview 
 
 
 
5.9.3  Factory Calibration — Overview 
Related Topics 
 •  Auto Self-Calibration — Overview (p. 89) 
 •  Auto Self-Calibration — Details (p. 337) 
 •  Auto Self-Calibration — Errors (p. 475) 
 •  Offset Voltage Compensation (p. 323) 
 •  Factory Calibration (p. 86) 
 •  Factory Calibration or Repair Procedure (p. 461) 
The CR800 uses an internal voltage reference to routinely calibrate itself. 
Campbell Scientific recommends factory recalibration as specified in 
Specifications 
(p. 91). If calibration services are required, see Assistance (p. 5). 
 
5.9.4  Internal Battery — Overview 
Related Topics: 
 •  Internal Battery — Quickstart (p. 38)   
 •  Internal Battery — Details
 (p. 457)   
 
Warning Misuse or improper installation of the internal lithium battery 
can cause severe injury. Fire, explosion, and severe burns can result. Do 
not recharge, disassemble, heat above 100 °C (212 °F), solder directly to 
the cell, incinerate, or expose contents to water. Dispose of spent lithium 
batteries properly. 
The CR800 contains a lithium battery that operates the clock and powers SRAM 
when the CR800 is not externally powered. Voltage of the battery is monitored 
from the CR800 Status table (LithiumBattery 
(p. 543)). Replace the battery as 
directed in Internal Battery — Details 
(p. 457). 
The lithium battery is not rechargeable. Its design is one of the safest available 
and uses lithium thionyl chloride technology. Maximum discharge current is 
limited to a few mA. It is protected from discharging excessive current to the 
internal circuits (there is no direct path outside) with a 100 ohm resistor. The 
design is UL listed. See: 
http://www.tadiran-batterie.de/download/eng/LBR06Eng.pdf. 
 
5.10  Datalogger Support Software — Overview 
Related Topics: 
 • Datalogger Support Software — Quickstart (p. 39)   
 • Datalogger Support Software — Overview
 (p. 86)   
 • Datalogger Support Software — Details
 (p. 396)   
 • Datalogger Support Software — Lists