EasyManuals Logo

Campbell CR800 Series User Manual

Campbell CR800 Series
598 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #221 background imageLoading...
Page #221 background image
Section 7. Installation
221
Calibration Report for Salinity Sensor
CRBasic Variable At Deployment At Seven-Day Service
SimulatedSalinitySignal
output
1350 mV
1345 mV
KnownSalintiy (standard
solution)
30 mg/l 30 mg/l
SalinityMultiplier
0.05 mg/l/mV 0.05 mg/l/mV
SalinityOffset
-37.50 mg/l -37.23 mg/l
Salinity reading 30 mg/l 30 mg/l
1. Send CRBasic example FieldCal() Offset (p. 221) to the CR800. A terminal
configured for excitation has been programmed to simulate a sensor output.
2. To simulate the salinity sensor in a simulated-calibration condition, (in the
field it would be placed in a 30 mg/l standard solution), place a jumper wire
between terminals VX1 and SE1. The following variables are preset by the
program: SimulatedSalinitySignal = 1350, KnownSalinity = 30.
3. To start a simulated calibration, set variable CalMode = 1. When CalMode
increments to 6, offset calibration is complete. The calibrated offset will equal
-37.48 mg/l.
4. To continue this example and simulate an offset-drift condition, set variable
SimulatedSalinitySignal = 1345.
5. To simulate seven-day-service calibration conditions (30 mg/l standard
solution), the variable KnownSalinity remains at 30.0. Change the value in
variable CalMode to 1 to start the calibration. When CalMode increments to 6,
the seven-day-service offset calibration is complete. Calibrated offset will
equal -37.23 mg/l.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Campbell CR800 Series and is the answer not in the manual?

Campbell CR800 Series Specifications

General IconGeneral
BrandCampbell
ModelCR800 Series
CategoryData Loggers
LanguageEnglish

Related product manuals