Section 7.  Installation 
 
 
 Calibration Report for Salinity Sensor 
CRBasic Variable  At Deployment  At Seven-Day Service 
SimulatedSalinitySignal 
output 
1350 mV 
1345 mV 
KnownSalintiy (standard 
solution) 
30 mg/l  30 mg/l 
SalinityMultiplier 
0.05 mg/l/mV  0.05 mg/l/mV 
SalinityOffset 
-37.50 mg/l  -37.23 mg/l 
Salinity reading  30 mg/l  30 mg/l 
 
1.  Send CRBasic example FieldCal() Offset (p. 221) to the CR800.    A terminal 
configured for excitation has been programmed to simulate a sensor output. 
2.  To simulate the salinity sensor in a simulated-calibration condition, (in the 
field it would be placed in a 30 mg/l standard solution), place a jumper wire 
between terminals VX1 and SE1.    The following variables are preset by the 
program: SimulatedSalinitySignal = 1350, KnownSalinity = 30. 
3.  To start a simulated calibration, set variable CalMode = 1. When CalMode 
increments to 6, offset calibration is complete. The calibrated offset will equal 
-37.48 mg/l. 
4.  To continue this example and simulate an offset-drift condition, set variable 
SimulatedSalinitySignal = 1345. 
5.  To simulate seven-day-service calibration conditions (30 mg/l standard 
solution), the variable KnownSalinity remains at 30.0.    Change the value in 
variable CalMode to 1 to start the calibration. When CalMode increments to 6, 
the seven-day-service offset calibration is complete. Calibrated offset will 
equal -37.23 mg/l.